AD111 pH-ORP-TEMP meter (ADWA) is a professional portable meter for pH, mV and temperature. pH to two decimals, automatic calibration.

ADWA AD111: a professional portable meter for pH, ORP (mV) and temperature
The AD111 pH-ORP-TEMP meter is a portable meter with an external electrode and temperature probe for field and laboratory work. It measures pH and sample temperature, and with a suitable ORP electrode also the redox potential in mV. Automatic temperature compensation adjusts the pH measurement to the sample temperature, the HOLD key freezes the reading, and an indicator warns of low batteries.
The set includes an A1230B pH electrode (epoxy body, BNC connector, 1 m cable), an A7662 stainless steel temperature probe (1 m cable), pH 4.01, 7.01 and 10.01 calibration solutions (20 mL each), four AA batteries and instructions. The ORP electrode is not included in the set; it is available separately as the A3230B ORP electrode.
According to the official ADWA manufacturer's instructions it measures pH in the range from −2.00 to 16.00 with a resolution of 0.01 pH, mV in the range ±1000 mV and temperature from −5.0 to 105.0 °C.
How it works and how to calibrate it
Dip the electrode and temperature probe into the sample, stir gently and wait for the value to settle. The SET/HOLD key selects the measuring range (pH, mV or °C), and a longer press freezes the reading. To measure mV, connect the ORP electrode and switch to mV. The mV range is factory calibrated, so you do not calibrate it yourself. You calibrate only pH.
pH calibration is automatic: the meter recognises the 4.01, 7.01 or 10.01 buffer by itself. Two-point calibration is recommended. Dip the electrode and probe into the 7.01 buffer, press CAL and confirm with SET/HOLD when the value stops flashing. Then dip them into the 4.01 or 10.01 buffer and confirm again. If the buffer value is wrong, the message Ec appears.
Who is it for
Choose it when besides pH you also need ORP and a meter with an external electrode. If you do not need ORP, the cheaper choice is the AD110 pH/temp meter. For a waterproof version see the AD131 or AD132, and for a simple pocket ORP and pH meter the AD14 pH/ORP meter. For frequent measurement of milk pH the page recommends the special A1131B glass electrode, which is bought separately.
Storing and cleaning the electrode
Keep the electrode in its protective bottle so that the glass bulb stays moist. Check the cable and electrode regularly: if you notice cracks or damaged insulation, replace the electrode, and keep the connectors clean and dry. Clean the electrode with a pH electrode cleaning solution and then recalibrate the meter. The ORP meter can be checked with the ORP control solution.
If you are looking for an AD111 pH-ORP-TEMP meter for measuring pH, ORP and temperature in the laboratory or in the field, this is a professional portable ADWA meter with an external electrode. Also see all ADWA pH and ORP meters.